COVID-19 RT-qPCR Detection Kit (RUO)
Size: 100 Rxns
In view of the joint global efforts of advancing collaborative research in diagnostics, therapeutics, and vaccination in the fight against the COVID-19 (SARS-CoV-2) pandemic, Bio-Helix has specifically developed the LifeDireX COVID-19 RT-qPCR Detection Kit for human respiratory tract specimens. The kit is characterized by: (1) High specificity for the RdRP and N target markers (2) Data obtained in less than 2 hours; and (3) Compatible with standard RT-qPCR machines (ABI 7500, Bio-Rad CFX96, QuantStudio’s 7 Flex).
Part No. |
Component |
Volume |
Reactions/Kit |
QP019-0100-1 |
2X RT-qPCR MasterMix |
1.25 ml |
100 |
QP019-0100-2 |
RT-qPCR Enzyme Mix |
40 μl |
100 |
QP019-0100-3 |
COVID-19 Primers/Probes |
200 μl |
100 |
QP019-0100-4 |
Positive Control |
100 μl |
20 |
QP019-0100-5 |
Negative Extraction Control |
1.0 ml |
20 |
QP019-0100-6 |
Nuclease-Free Water |
1.0 ml |
100 |
Application
- Gene Expression (mRNA) Analysis》
- Copy Number Analysis
- SNP Genotype Analysis
Important notes
- Shake gently before use to avoid foaming and low-speed centrifugation.
- Reduce the exposure time.
- During operation, always wear a lab coat, disposable gloves, and protective equipment.
Q: Which genes does the Kit target?
A: It covers 3 genes: N, RdRP, and RP genes
Q: What are the shipment measurements?
A: We want to make sure our kits get to the destination at a cooling temperature, therefore we use Dry Ice and Ice packs in combination.
Q: Is the kit CE or FDA accredited?
A: This kit is for Research Use only.
Q: Can we run the kit in any equipment?
A: Our kit is compatible with RT-qPCR machine standards and our selected one is Bio-Rad CFX96. Please let us know which machine you are using. You are welcome to try the machine of your choice.
- PCR Reaction: Thaw and assemble the following components in a 0.2 ml PCR tube on ice just prior to use: COVID-19 Primers, COVID-19 Probes, 2X RT-qPCR MasterMix, and RT- qPCR Enzyme Mix. Caution: Do not add more than one RNA sample into a single qPCR tube. Mix gently. If necessary, centrifuge briefly.
Component | 20 μl Patient Sample | 20 μl Positive Control | 20 μl Positive Control | Negative Control |
RNA Sample | 5 μl | 0 μl | 0 μl | 0 μl |
COVID-19 Primers/Probes | 2 μl | 2 μl | 2 μl | 2 μl |
2X RT-qPCR MasterMix | 10 μl | 10 μl | 10 μl | 10 μl |
RT-qPCR Enzyme Mix | 0.4 μl | 0.4 μl | 0.4 μl | 0.4 μl |
RT-qPCR Enzyme Mix | 0 μl | 5 μl | 0 μl | 0 μl |
Negative Extraction Control | 0 μl | 0 μl | 5 μl | 0 μl |
Nuclease – Free H2O | 2.6 μl | 2.6 μl | 2.6 μl | 7.6 μl |
- Use the Nuclease-free H2O for the Negative Control while using Positive Control for the Positive Control setup. Cap tubes and place in the thermal cycler.
- Process in the thermal cycler for 42 cycles as follows:
Steps | Temperature/Time | Cycle |
cDNA Synthesis | 15 minutes at 42°C | 1 |
Pre-Denaturation | 10 minutes at 95°C | 1 |
Denaturation | 15 seconds at 95°C | 40 |
Annealing | 60 seconds at 60°C | |
Melting curve | Refer to specific guidelines for instrument used |
Note: Optimal conditions for amplification will vary depending on the primers and thermal cycler used. It may be necessary to optimize the system for individual primers, template, and thermal cycler.
- Detection: As three channels (FAM, ROX, HEX) are used in this one tube qPCR assay, we recommend to perform the channel calibration as requested by its manufacturer. Please refer to the instrument’s user manual to perform this calibration. Choose the FAM, ROX, and HEX channels for each sample to be tested with the LifeDireX COVID-19 RT-qPCR Detection Kit. Select “None” for ROX passive reference on any qPCR machine requiring ROX as the reference dye.
Expected Performance of Controls
Control Type | Used to Monitor | Expected Results and Ct Values | ||
N (FAM) | RP (HEX) | RdRP (ROX) | ||
Positive | Flawed assay setup and reagent failure, including degraded primer and probe |
Positive Ct < 40.0 |
Negative Ct ND |
Positive Ct < 40.0 |
Negative (“NTC”) |
Assay or extraction reagent contamination |
Negative Ct ND |
Negative Ct ND |
Negative Ct ND |
Negative Extraction Control |
Cross-contamination |
Negative Ct ND |
Positive Ct < 40.0 |
Negative Ct ND |
ND = Not Dtected.
Results are considered invalid if any control does not perform as specified above.
Interpretation of Results
SARS-CoV-2 | Interpretation | ||
N | RdRP | RP | |
+ | + | +/- | Positive |
If only one of the two targets is positive | +/- | Inconclusive Result | |
- | + | Negative | |
- | - | - | Negative |
Note: The results from this kit should be interpreted in combination with all relevant laboratory findings.
Name | Download |
---|---|
Safety Data Sheet|SDS | COVID-19_RT-qPCR_Detection_Kit_QP019_SDS.pdf |
Comparison Data | QP019-0100_Comparison_Data_of_COVID-19_Detection_Kit.pdf |
PROTOCOL | QP019_Protocol_v3_20210305.pdf |